Model |
TH2838 |
Test Signal Source |
Output impedance |
100Ω, ±1% @1kHz |
Frequency |
Range |
20Hz-2MHz |
Resolution |
20.0000Hz - 99.9999Hz 0.1mHz |
100.000Hz - 999.999Hz 1mHz |
1.00000kHz - 9.99999kHz 10mHz |
10.0000kHz - 99.9999kHz 0.1Hz |
100.000kHz - 999.999kHz 1Hz |
1.00000MHz - 2.00000MHz 10Hz |
AC test signal |
Rated value(ALC OFF):Set the voltage as the Hcur voltage when the test terminal is open Set the current as the Hcur current when the test terminal is short Constant value(ALC ON):Keep the voltage in DUT is the same as the set valueKeep the current in DUT is the same as the set value |
AC signal |
Voltage range |
5mVrms -- 2Vrms |
Resolution |
5mVrms -- 0.2Vrms 100μVrms |
0.2Vrms -- 0.5Vrms 200μVrms |
0.5Vrms -- 1Vrms 500μVrms |
1Vrms -- 2Vrms 1mVrms |
2Vrms -- 5Vrms 2mVrms |
5Vrms -- 10Vrms 5mVrms |
10Vrms -- 20Vrms 10mVrms |
Current range |
50μArms -- 20mArms |
Resolution |
50μArms -- 2mArms 1 μArms |
2mArms -- 5mArms 2 μArms |
5mArms -- 10mArms 5 μArms |
10mArms -- 20mArms 10μArms |
20mArms -- 50mArms 20μArms |
50mArms--100mArms 50μArms |
Rdc test |
Voltage range |
100mV - 2V |
Resolution |
100μV |
Current range |
0mA- 20mA |
Resolution |
1μA |
DC Bias |
Voltage range |
0V - ± 10V |
Resolution |
0V -- 5V 100μV |
5V -- 10V 1mV |
Current range |
0mA- ± 100mA |
Resolution |
0 A -- 50mA 1μA |
50mA -- 100mA 10μA |
Display |
Size/Type |
7-inch (diagonal)TFT LCD display |
Proportion |
16:9 |
Resolution |
800×RGB×480 |
Test function |
Test parameter |
Cp-D,Cp-Q,Cp-G,Cp-RpCs-D,Cs-Q,Cs-RsLp-D, Lp-Q, Lp-G, Lp-Rp, Lp-RdcLs-D, Ls-Q, Ls-Rs, Ls-Rdc, RdcR-X, Z-θd, Z-θrG-B, Y-θd, Y-θrVdc-Idc |
Mathematics function |
A(X+B)+C, X is test parameter, A, B,C is input parameter |
Equivalent circuit |
Series, parallel |
Deviation measurement |
Absolute deviation Δ compared with the nominal valuePercentage deviation Δ% compared with the nominal value |
Calibration function |
OPEN, SHORT, LOAD |
Range selection |
AUTO, HOLD |
Range |
LCR |
100mΩ, 1Ω, 10Ω, 20Ω, 50Ω, 100Ω, 200Ω, 500Ω, 1kΩ, 2kΩ, 5kΩ, 10kΩ, 20kΩ, 50kΩ, 100kΩ, total 15 ranges |
Rdc |
1Ω, 10Ω, 20Ω, 50Ω, 100Ω, 200Ω, 500Ω, 1kΩ, 2kΩ, 5kΩ, 10kΩ, 20kΩ, 50kΩ, 100kΩ, total 15 ranges |
Trigger mode |
INT, MAN, EXT, BUS |
Trigger delay |
0 s -- 999 s, resolution 100us |
Test terminal configuration |
Four-pair |
Test cable length |
0m, 1m |
Test average |
1-255 times |
Test time(ms) |
Speed mode |
20Hz |
100Hz |
1kHz |
10kHz |
100kHz |
1MHz |
2MHz |
FAST |
330 |
100 |
20 |
7.7 |
5.7 |
5.6 |
5.6 |
MED |
380 |
180 |
110 |
92 |
89 |
88 |
88 |
LONG |
480 |
300 |
240 |
230 |
220 |
220 |
220 |
Test display range a 1×10-18 ; E 1×1018 |
Cs, Cp |
±1.000000 aF -- 999.9999 EF |
Ls,Lp |
±1.000000 aH -- 999.9999 EH |
D |
±0.000001 -- 9.999999 |
Q |
±0.01 -- 99999.99 |
R, Rs, Rp, X, Z, Rdc |
±1.000000 aΩ -- 999.9999 EΩ |
G,B,Y |
±1. 000000 aS -- 999.9999 ES |
Vdc |
±1.000000 aV -- 999.9999 EV |
Idc |
±1.000000 aA -- 999.9999 EA |
θ r |
±1.000000 a rad -- 3.141593 rad |
θ d |
±0.0001 deg -- 180.0000 deg |
Δ% |
±0.0001% -- 999.9999% |
t |
-99.99°C -- 1000.00°C |
Turn Ratio (extension pending) |
±0.000000 -- 1000.000 |
Basic test accuracy |
0.05% (the details refer to the instruction) |
List sweep |
Sweep points |
Up to 201 points |
Sweep Parameters |
Test frequency, AC voltage, AC current, DC BIAS voltage, DC BIAS current |
Trigger mode |
SEQ |
Once triggered, test at the sweep points. /EOM/INDEX will be output one time. |
STEP |
Once triggered, test at one sweep point. /EOM/INDEX will be output at each point, but the list sweep comparator results only be output at the last /EOM. |